Dynamic Digital Radiography/Fluoroscopy System (DR,65KW)
Model: SHO-DDR01 System
Fill in the form to get more product information
Product Introduction
SHO-DDR01 Digital Gastrointestinal MachineIntroduction
It is Multi-purpose R/F System is a flexible and compact system with various functions such as radiography, fluoroscopy, stomach examination,gastrointestinal tract examination, etc. Its wide-range movements and auto DR operations reduce the patient movements and make the operation easier and faster.
Fluoroscopy: Dynamic a-Si flat panel detector is designed for real-time imaging. The 17×17-inch image sensor has direct deposition CsI , which provides excellent low-dose imaging at frame rates up to 30 fps. It is a digital fluoroscopic X-ray system, which can be used in both general radiography, fluoroscopy for gastrointestinal, urolog applications.
Radiography: Retractable tube arm, max SID is 1800mm. It could provide digital radiography of various body parts without blind area.
Features
◆High-performance and high-capacity with big focus and high speed, which is suitable for long-time diagnosis.
◆High inverting frequency and high voltage generator ensures output of constant DC.
◆Dynamic flat panel detector brings digital radiography and fluoroscopy of various positions directly.
◆ Strong and weak electricity separation and excellent EMC characteristics.
◆High density grid further improves X-ray quality and provides images with good effect.
◆Designed for low-dose fluoroscopic imaging
Specifications
Ambient Conditions
Ambient temperature | 10℃~40℃ |
Relative humidity | 30%~75% |
Atmospheric | 70~106kPa |
X-ray Generator
Model | SHO-DDR01 Series | |||
Output power | 32kw | 50kw | 65kw | 80kw |
Current | 10-400mA | 10-630mA | 10-800mA | 10-1000mA |
Voltage | 40kv-150kv | 40kv-150kv | 40kv-150kv | 40kv-150kv |
mAs | 0.1-400mAs | 0.1-630mAs | 0.1-800mAs | 0.1-1000mAs |
Exposure time | 1ms~6300ms | |||
Focus size | 1.0/2.0 | 0.6/1.2 | 0.6/1.2 | 0.6/1.2 |
X-ray Tube
Anode type | Rotating Anode |
Focal Spot value | 0.6mm /1.2mm |
Maximum voltage | 150KV |
Revolving speed of anode | 2800rpm |
Inherent filtration | 1.0mmAl |
Anode heat storage capacity | 300kHU |
Maximum cooling rate of anode | 870W |
Housing heat storage capacity | 900kJ |
X-ray detector(Iray brand Flat Panel Detector)
Type | Amorphous Silicon |
Scintillator | Cesium Iodide |
Active Area | 17x17Inch(43cm*43cm) |
Active Pixel | 3072 × 3072 |
Pixel Pitch | 139um |
Limiting resolution | 36Lp/cm |
Data acquisition time | ≤2s |
Collimator
Type | Electric and multileaf collimator |
Power | 150W; 24VAC 1 |
Lamp timer | Automatic illumination with timer for lamp(30S) |
Filtration | 1.2mmAL |
Mobile patient table
Type | Table for gastrointestinal imaging |
Tabletop lateral movement | 240mm(-120mm~+120mm) |
Rotating range | -20°~+90° |
Mechanical movement
SID | 1000mm~1800mm |
Detector Cassette movement | 1100mm |
Workstation
Monitor | 24’’ LCD AOC brand |
CPU | ≥2.0GHz |
Memory | ≥2GB |
HDD | ≥500GB |
DICOM3.0 | Query for integration with any PACS |
Functions | Import/export function |
Image info | |
Management of patient info | |
Post processing | |
Measurement etc. |